Description
PAMAS SVSS – Laboratory Particle Counter for Water-Based and Pharmaceutical Liquids
PAMAS SVSS is a high-precision particle counter designed for laboratory analysis of water-based liquids, pharmaceutical products and other low-viscosity media. The system is used for contamination control, particle size analysis and component cleanliness monitoring within the pharmaceutical industry, research and quality control laboratories.
The instrument can be used either as a standalone system or together with advanced analysis software. A stepper motor-controlled precision syringe transports the sample through the laser sensor at an optimal flow rate, ensuring highly accurate and reproducible measurement results.
Thanks to interchangeable syringes, sample volumes ranging from only 100 µl up to 1,000 ml can be analysed using the same system. A dedicated Small Volume Kit is also available for very small sample volumes.
Benefits
Designed for pharmaceutical and water-based liquids
Very high measurement accuracy and reproducibility
Automatic control of sample volume and flow rate
Sample volumes from 100 µl to 1,000 ml
16 freely definable particle size channels
32 channels available as an option
Integrated magnetic stirrer
Compatible with different laser sensors
Can be used with a wide range of sample containers
Ideal for pharmaceutical and quality control applications
Typical Applications
• Parenteral pharmaceuticals according to international pharmacopoeias
• Infusion solutions
• Ophthalmic liquids
• Oral pharmaceutical solutions
• Water and purified water
• Suspensions
• Cleaning solutions
• Quality control within the pharmaceutical industry
• Laboratory analysis and research
Technical Information
Sample volume: 100 µl–1,000 ml
Flow control: Stepper motor-controlled precision syringe
Syringe sizes: 1 ml, 10 ml and 25 ml
Particle channels: 16 standard, 32 optional
Magnetic stirrer: Integrated
Operation: Standalone or software-controlled
Sensor Options
PAMAS HCB-LD-50/50
Measurement range:
• 1–200 µm according to ISO 21501-3
• 1–400 µm according to ISO 21501-3 (optional)
Maximum particle concentration:
• 24,000 particles/ml at 10 ml/min
PAMAS HCB-LD-25/25
Measurement range:
• 1–200 µm according to ISO 21501-3
Maximum particle concentration:
• 120,000 particles/ml at 10 ml/min
PAMAS SLS-25/25
Measurement range:
• 0.5–20 µm according to ISO 21501-2
Maximum particle concentration:
• 13,000 particles/ml at 10 ml/min
Standards
PAMAS SVSS complies with or supports, among others:
USP <787>
USP <788>
USP <789>
ISO 21501-2
ISO 21501-3
Ph. Eur. 2.9.19
Ph. Eur. 2.9.53
BP XIII
JP 6.0.7
KP XII
IP 2.2.32
IP 2.5.9
CE Marked